| Model Number |
TFGT-25A |
| Medical Device Problem Code |
Device Stenosis (4066)
|
| Health Effect - Clinical Code |
Aortic Valve Stenosis (1717)
|
| Date of Event |
05/15/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
It was reported valve was severely scarred and difficult to remove around 6 years of trifecta valve implant.A more comprehensive assessment, including histopathological examination of the valve tissue could not be performed as the device was not returned for analysis.However, based on the information received, the reported stenosis is consistent with structural valve deterioration (svd), specifically fibro-calcific svd (fcsvd), which is a well-known complication of tissue heart valve replacement surgery.A variety of factors may contribute to svd, including patient, biological, and implant related factors: no implant related factors could be confirmed from the information received from the field as information related to implant procedure was not provided.Biological factors which can result in stenosis such as calcification (from patient conditions predisposing to elevated calcium like renal disease etc.), immobilizing thrombus, or pannus formation reducing the valve diameter also could not be confirmed as the valve was not returned for histopathological examination.The exact cause of the reported incident could not be conclusively determined.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2018, a 25mm trifecta gt valve was implanted in the patient.On (b)(6) 2024, it was found that the valve was severely scarred and difficult to remove.The valve was removed and a new 25mm epic valve was implanted.The patient was reported stable.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2018, a 25mm trifecta gt valve was implanted in the patient.On (b)(6) 2024, it was found that the valve was severely scarred and difficult to remove.The valve was removed and a new 25mm epic valve was implanted.The patient was reported stable.No further information was available.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of valve being severely scarred and difficult to remove was reported.The trifecta valve was explanted after over 5 years of implant.A more comprehensive assessment, including histopathological examination of the valve tissue could not be performed as the device was not returned for analysis.The device history record was also reviewed to ensure that each manufacturing and inspection operation was performed, and the product met all specifications.Based on the information received, the cause of the reported incident could not be conclusively determined.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.This individual case was received by the manufacturer as part of a bulk report and is being submitted as a separate medical device report (mdr) specific to the identified device and patient.
|
| |
|