| Model Number |
CF-H190I |
| Medical Device Problem Code |
Failure to Clean Adequately (4048)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The evaluation of the event is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was observed during the device inspection, that the colonovideoscope exhibited foreign material in jet tube.There were no reports of patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and device evaluation.Correction e1: establishment name was inadvertently left blank.Additional information: d4 (udi), d8, h3, h6.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, olympus confirmed foreign material came out of the device, but the type of the material could not be identified.There was no physical damage at the location where the foreign material remained.A definitive root cause cannot be determined.The event can be detected/prevented by following the instructions for use which state: ifu states the detection method in gif/cf/pcf-190 series operation manual chapter 3 preparation and inspection.Ifu states the preventive measure in gif/cf/pcf-190 series reprocessing manual chapter 5 reprocessing the endoscope.Olympus will continue to monitor field performance for this device.
|
| |
|