C.R. BARD, INC. (BASD) -3006260740 POWERPICC CATHETER WITH SHERLOCK 3CG (TPS) STYLET 4F; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
|
Back to Search Results |
|
| Model Number |
N/A |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/26/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The device has not been returned to the manufacturer for evaluation.
|
| |
|
Event or Problem Description
|
|
It was reported that rn states picc is leaking.Upon assessment picc would not flush easily, and when i tried to draw blood, air came into syringe.I immediately removed the dressing and upon flushing i saw a stream of fluid coming from the place in the image.Its not a slit, but a pinhole.I dc ed picc intact and planning on placing new picc later today.No patient injury.No other information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The complaint of a leak was confirmed.The product returned for evaluation was one 4 fr sl powerpicc catheter.The investigation findings are consistent with damage accumulated through flexural fatigue.Flexural fatigue occurs due to cyclic kinking of the catheter tube in which physiological, placement, usage, and mechanical factors may gradually form a crack(s) in the catheter.The returned product sample was evaluated and a split was observed in the catheter tubing just distal of the molded joint.The catheter split contained physical features associated with material fatigue, and the characteristics observed which supported this type of failure included: ¿ damage which was circumferentially aligned.¿ fracture edges which were rounded and polished due to repeated material wear.Repetitive kinking of the indwelling catheter appears to have contributed to the observed leak; however, the specific circumstances surrounding how kinking developed into a fracture within the catheter tubing were ultimately unknown.A measurement of the catheter outer diameter, inner diameter and wall thickness were taken.The catheter was found to be within specification.This complaint will be recorded for future trending and monitoring purposes.
|
| |
|
Event or Problem Description
|
|
It was reported that rn states picc is leaking.Upon assessment picc would not flush easily, and when i tried to draw blood, air came into syringe.I immediately removed the dressing and upon flushing i saw a stream of fluid coming from the place in the image.Its not a slit, but a pinhole.I dc ed picc intact and planning on placing new picc later today.No patient injury.No other information was provided.
|
| |
|
Search Alerts/Recalls
|
|
|