STRYKER ENDOSCOPY-SAN JOSE DOUBLE BITE SHAVER BLADE S-SERIES 4.0MM EACH - ORDER MULTIPLES OF 5; BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL Back to Search Results
Catalog Number 0375543000
Medical Device Problem Code Delivered as Unsterile Product (1421)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/13/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that there was a breach in the sterile barrier.
 
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.
 
Event or Problem Description
It was reported that there was a breach in the sterile barrier.
 
Additional Manufacturer Narrative
Alleged failure: the packaging of the shaver blades was already open on one side when i took them out of the box.The failure(s) identified in the investigation is consistent with the complaint record.Probable root causes could be an inadequate adhesive application process, inadequate positioning into the primary package, shipping and handling damage, or improper environmental conditions.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DOUBLE BITE SHAVER BLADE S-SERIES 4.0MM EACH - ORDER MULTIPLES OF 5
Common Device Name
BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
lucas wolski
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key20488095
Report Number0002936485-2024-00709
Device Sequence Number17024026
Product Code GFA
UDI-Device Identifier37613327061759
UDI-Public37613327061759
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 02/26/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number0375543000
Device Lot Number23282CE2
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/27/2024
Supplement Date Received by Manufacturer09/27/2024
Initial Report FDA Received Date10/18/2024
Supplement Report FDA Received Date02/26/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/09/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-