• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WELCH ALLYN INC. PRO 6000 W/SMALL CRADLE; THERMOMETER, ELECTRONIC, CLINICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WELCH ALLYN INC. PRO 6000 W/SMALL CRADLE; THERMOMETER, ELECTRONIC, CLINICAL Back to Search Results
Model Number 06000-200
Medical Device Problem Code Excessive Heating (4030)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/10/2024
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported that the probe tip of their braun pro6000 overheated.There were no allegations of injury.This event was captured under hillrom complaint record#: (b)(4).
 
Additional Manufacturer Narrative
The customer reported that the probe tip of their braun pro6000 overheated.There were no allegations of injury.The braun/welch allyn thermoscan pro 6000 ear thermometer is indicated for the intermittent measurement of human body temperature for patients having ages ranging from normal weight (full term) newborn to geriatric adults in a professional use environment.The probe cover is used as a sanitary barrier between the infrared thermometer and the ear canal.To help ensure accurate temperature measurements, the sensor itself is warmed to a temperature close to that of the human body.When the braun/welch allyn thermoscan is placed in the ear, it continuously monitors the infrared energy until a temperature equilibrium has been reached and an accurate measurement can be taken.Hillrom/welch allyn¿s investigation of pro 6000 hot tip complaints confirmed that aggressive cleaning and decontamination practices can cause liquid ingress.Consequently, liquid ingress can adversely affect the temperature sensor causing the pro 6000 device to behave inconsistently and to overheat the speculum tip.It is believed that replication of the malfunction with the returned devices from the customer at the manufacturing site has not been possible as any fluid that may have ingressed had likely evaporated, therefore not showing the hot tip malfunction.Based on hillrom¿s ability to replicate the malfunction of a hot tip on new devices that can potentially go above the built in risk mitigation's of a safety cut off that could potentially cause a more serious injury we have deemed this complaint to be reportable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRO 6000 W/SMALL CRADLE
Common Device Name
THERMOMETER, ELECTRONIC, CLINICAL
Manufacturer (Section D)
WELCH ALLYN INC.
4341 state street
skaneateles falls 13153
Manufacturer Contact
frances coote
1069 state route 46 east
batesville, IN 47006
8129310130
MDR Report Key20490111
Report Number1316463-2024-00137
Device Sequence Number4096368
Product Code FLL
UDI-Device Identifier00732094309003
UDI-Public732094309003
Combination Product (Y/N)N
Initial Reporter CountryFR
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 10/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number06000-200
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/10/2024
Initial Report FDA Received Date10/20/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-