• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL LLC ARROW CANNON II PLUS REPLACEMENT HUB SET; KIT, REPAIR, CATHETER, HEMODI

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL LLC ARROW CANNON II PLUS REPLACEMENT HUB SET; KIT, REPAIR, CATHETER, HEMODI Back to Search Results
Catalog Number CAR-02400
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/25/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "on (b)(6) 2024, the doctor found the luer hub/fiting has a crack during use on the patient." there was no medical intervention required.The patient's current condition is reported as "fine".
 
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that "on (b)(6) 2024, the doctor found the luer hub/fiting has a crack during use on the patient." the operator changed to a new set.There was no medical intervention required.The patient's current condition is reported as "fine".
 
Additional Manufacturer Narrative
Qn# (b)(4).The customer returned one connector assembly from a cannon ii plus replacement hub kit for analysis.Signs of use in the form of biological material were observed on and within the device.Visual analysis revealed scratch/stress markings on both the blue venous luer hub and red arterial luer hub.The appearance of this damage is consistent with overtightening or repeated tightening of the hubs.Both luer connections were functionally tested per the product instructions-for-use (ifu).The ifu provided with this kit instructs the user, "flush each catheter lumen with sterile normal saline for injection, to establish patency and prime lumen(s) and flush the irrigation tube." a lab inventory syringe filled with water was used to flush each extension line.No leaks were observed from the red arterial luer hub; however, leaking was observed from the crack in the blue venous luer hub.A manual tug test confirmed both luer hub were securely attached to their respective lumens.A device history record review was performed, and no relevant findings were identified.The ifu warns the user, "examine catheter and extension lines before and after each treatment for any signs of damage.Repeated over tightening of bloodlines, syringes and caps will reduce connector life and could lead to potential connector failure." the report of cracked luer hubs was confirmed through complaint investigation of the returned sample.Visual analysis revealed both luer hubs contained a crack and damage consistent with overtightening or repeated tightening of the hubs.A device history record review was performed, and no relevant findings were identified.Based on the customer report and the sample received, unintentional use error likely caused or contributed to this event.Teleflex will continue to monitor and trend for reports of this nature.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW CANNON II PLUS REPLACEMENT HUB SET
Common Device Name
KIT, REPAIR, CATHETER, HEMODI
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL DE MEXICO S.A. DE C.V.
ave. washington 3701
colonia panamericana, chihuahua
chihuahua 31200
MX   31200
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key20491063
Report Number9680794-2024-01037
Device Sequence Number4106221
Product Code NFK
UDI-Device Identifier00801902096111
UDI-Public(01)00801902096111(11)230802(17)260630(10)33F23G0557(21)
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K020430
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2026
Device Catalogue NumberCAR-02400
Device Lot Number33F23G0557
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/17/2024
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/26/2024
Supplement Date Received by Manufacturer11/06/2024
Initial Report FDA Received Date10/21/2024
Supplement Report FDA Received Date11/06/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/02/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NONE REPORTED.; NONE REPORTED.
Patient SexUnknown
-
-