| Catalog Number |
201001120 |
| Medical Device Problem Codes |
Break (1069); Naturally Worn (2988); No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/15/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the actis broaches are damaged and require replacements.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Additional information provided the instruments are not broken.Coding updated the impacted products from broken to worn.Depuy synthes joint reconstruction does not consider this failure to be a reportable malfunction at this time.There is no indication that any patient consequence or injury occurred either.Further updates will only be provided if additional information is received that changes the regulatory determination.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #(b)(4).Investigation summary: according to the information provided, it was reported that swap out old actis broaches with new ones.There was no pieces broke off and new broaches have extra slot for extraction.The product was returned to j&j medtech orthopaedics for evaluation.Visual analysis revealed that the post of the actis broach sz 12 shows scratches and signs of heavy use.No other defects that could impact the assembly of the broach were observed.A functional test was performed with a mating device sample revealed that the broach is able to be fully assembled.A dimensional inspection was not performed as it is not applicable to the complaint condition.Information provided has been reviewed and determined that there is no allegation of deficiency related to the identity, quality, durability, reliability, safety, effectiveness or performance of the j&j medtech orthopaedics device.However, due to the findings of the actis broach, the overall complaint was confirmed as the observed condition would have contributed to a device issue.Based on the investigation findings, the potential cause is traced to end of life, and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
| |
|
Search Alerts/Recalls
|