• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CHRISTOPH MIETHKE GMBH & CO KG PROGAV 2.0 WITH SHUNTASSISTANT 25; HYDROCEPHALUS MANAGEMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CHRISTOPH MIETHKE GMBH & CO KG PROGAV 2.0 WITH SHUNTASSISTANT 25; HYDROCEPHALUS MANAGEMENT Back to Search Results
Model Number FX414T
Medical Device Problem Code Infusion or Flow Problem (2964)
Health Effect - Clinical Code Hydrocephalus (3272)
Date of Event 09/30/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Visual inspection: during the investigation, scratches on the outer housing of the valves, but no significant deformations or damage were determined.Permeability test: a permeability test has shown that all components are permeable.Computer controlled test: to investigate the claim of over-drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valves are tested in both the horizontal as well as the vertical positions.The results show that the valves do not operate within the permissible tolerance in their respective relevant positions an accelerated outflow of both valves could be determined.Adjustment test: the progav 2.0 was tested and is adjustable to all specified pressures.Braking force and brake function test: the brake functionality test has shown that the brake function operates as expected and the braking force is within the given tolerances.Internal inspection: after dismantling of the valves, deposits were found in both valves.Results: based on the investigation results, we can determine an accelerated outflow at both valves.The deposits visible in the valves have led to the malfunction mentioned.Deposits caused by natural substances in the body, such as protein, blood or tissue particles, are among the known and unavoidable risks and side effects of hydrocephalus therapy.Even small amounts of proteins can impair the integrity of the valve.We can exclude a defect at the time of release.The shunt system met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
 
Event or Problem Description
It was reported that a progav 2.0 shunt system (#fx414t) was implanted.According to the complainant, the shunt system showed an over-drainage.The patient underwent a revision procedure performed on (b)(6) 2024.The complainant device has returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGAV 2.0 WITH SHUNTASSISTANT 25
Common Device Name
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
Manufacturer (Section G)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM   14469
Manufacturer Contact
joerg knebel
ulanenweg 2
potsdam, 14469
GM   14469
MDR Report Key20493461
Report Number3004721439-2024-00333
Device Sequence Number14505256
Product Code JXG
UDI-Device Identifier04041906136567
UDI-Public4041906136567
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K161853
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 10/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/18/2020
Device Model NumberFX414T
Device Catalogue NumberFX414T
Device Lot Number20028486
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/07/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/07/2024
Initial Report FDA Received Date10/21/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/19/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age15 YR
Patient SexMale
Patient Weight58 KG
-
-