U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. AIRWAY MOBILESCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. AIRWAY MOBILESCOPE Back to Search Results
Model Number MAF-DM2
Medical Device Problem Code Failure to Clean Adequately (4048)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that foreign material was observed in the operating part forceps channel of the mobilescope.There were no reports of patient involvement.
 
Event or Problem Description
It was observed during the device inspection, that the mobilescope exhibited foreign material in the operation unit forceps channel (brush insertion) section.There were no reports of patient involvement.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacture's (lm) review of the customer cleaning disinfection and sterilization (cds) processes, and the lms final investigation.Corrected field(s): b5, h3 and e1 (state, province, or territory - should be blank).New information added to the following field: h6.The device was returned for evaluation where the reported event was identified.Based on the results of the investigation, the foreign material could not be specified and a definitive root cause cannot be identified.Physical damage was found at the foreign material residue points and from the information obtained, it is unclear whether the reprocessing was carried out in accordance with the ifu, so the cause of the foreign material remaining could not be determined.A review of the device history record found no deviations that could have caused or contributed to the reported issue.The subject event can be detected and prevented by implementation in accordance with the below ifu.The instruction manual operation manual ¿maf-dm2/gm2/tm2 series, chapter 3 preparation and inspection¿ describes the methods for detection.The instruction manual reprocessing manual ¿maf-dm2/gm2/tm2 series, chapter 5 reprocessing the endoscope (and related reprocessing accessories)¿ describes the methods for prevention.Olympus will continue to monitor the field performance of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIRWAY MOBILESCOPE
Common Device Name
AIRWAY MOBILESCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key20493675
Report Number9610595-2024-21447
Device Sequence Number3615292
Product Code EOQ
UDI-Device Identifier04953170407260
UDI-Public04953170407260
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K201300
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 11/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberMAF-DM2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/11/2024
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/16/2024
Supplement Date Received by Manufacturer11/19/2024
Initial Report FDA Received Date10/21/2024
Supplement Report FDA Received Date11/25/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/23/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-