| Catalog Number |
R51004004L |
| Medical Device Problem Code |
Burst Container or Vessel (1074)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/19/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information is pending and will be submitted within 30 days upon receipt.
|
| |
|
Event or Problem Description
|
|
As reported, the balloon of a 4mm x 4cm saberx radianz percutaneous transluminal angioplasty (pta) balloon catheter ruptured during an inflation.There were no reported injuries to the patient.This was during a procedure to treat a severely calcified lesion.A 6f 110cm brite tip radianz catheter sheath introducer (csi) was used for this procedure as well as a vassallo guidewire.Additional information was requested but was not provided.The device will not be returned for evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint conclusion: as reported, the balloon of a 4mm x 4cm saberx radianz percutaneous transluminal angioplasty (pta) balloon catheter ruptured during an inflation.There were no reported injuries to the patient.This was during a procedure to treat a severely calcified lesion.A 6f 110cm brite tip radianz catheter sheath introducer (csi) was used for this procedure as well as a vassallo guidewire.Additional information was requested but was not provided.The reported ¿balloon burst¿ could not be confirmed as the device was not returned for analysis.The exact cause could not be determined.The reported severe calcification may have contributed to the reported event; it is likely to have caused damage to the balloon material that led to its rupture.However, without return of the product for analysis or films of the event it is difficult to draw a clinical conclusion between the device and the reported event.According to the ¿warnings¿ in the safety information in the instructions for use, ¿prior to angioplasty, the catheter should be examined to verify functionality and integrity, and ensure that its size and shape are suitable for the specific procedure for which it is to be used.Do not use if product damage is suspected or evident.To reduce the potential for vessel damage or the risk of dislodgement of particles it is very important that the inflated diameter of the balloon should approximate the diameter of the vessel just proximal and distal to the lesion.The balloon dimensions are printed on the product label.The compliance table incorporated with the product shows how balloon diameter increases as pressure increases.Do not exceed the rated burst pressure recommended on the label.The rated burst pressure is based on the results of in vitro testing.At least (b)(4) of the balloons (with a (b)(4) confidence) will not burst at or below their rated burst pressure.Use of a pressure monitoring device is recommended to prevent over-pressurization.Pressure in excess of the rated burst pressure can cause balloon rupture and potential inability to withdraw the catheter through the introducer sheath.Balloon rupture can cause vessel damage and the need for additional intervention.Use only the recommended balloon inflation medium (a 50/50 mixture by volume of contrast medium and normal saline).Never use air or any gaseous medium to inflate the balloon.¿ the information available does not suggest a design or manufacturing related cause for the reported event.Therefore, no corrective or preventive action will be taken at this time.
|
| |
|
Search Alerts/Recalls
|