• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN LP LLC NORTH HAVEN TRI-STAPLE 2.0; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN LP LLC NORTH HAVEN TRI-STAPLE 2.0; STAPLE, IMPLANTABLE Back to Search Results
Model Number SIGRADMT
Medical Device Problem Code Failure to Fire (2610)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/09/2024
Type of Reportable Event Malfunction
Event or Problem Description
According to the reporter, during a pulmonary resection, when the product was attached to the resection for removal of the pulmonary parenchyma, it could be clamped but the firing could not be done.Another lot of radial reload was used to resolve the issue.There was no patient injury.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that the reload was pre-fired and the interlock was engaged.The jaw of the reload was open.The sled was slightly advanced.It was reported that the instrument would not fire.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: failure to completely fire the reload will result in an incomplete cut and/or incomplete staple formation, which may result in poor hemostasis and/or leakage.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
According to the reporter, during a pulmonary resection, when the product was attached to the resection for removal of the pulmonary parenchyma, it could be clamped but the firing could not be done.Another lot of radial reload was used to resolve the issue.It was noted that the same handle was used to the new reload.There was no patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRI-STAPLE 2.0
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
Manufacturer (Section G)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key20494442
Report Number1219930-2024-05005
Device Sequence Number11814293
Product Code GDW
UDI-Device Identifier10884521543768
UDI-Public10884521543768
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K160176
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/14/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSIGRADMT
Device Catalogue NumberSIGRADMT
Device Lot NumberN3K2298Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/15/2024
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/09/2024
Supplement Date Received by Manufacturer12/26/2024
Initial Report FDA Received Date10/21/2024
Supplement Report FDA Received Date01/14/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/09/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-