| Model Number |
DRIVER SHAFT, T6, SELF RETAINING, AO |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/27/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Event or Problem Description
|
|
On 09/27/2024 it was reported by a sales representative via (b)(4).That (qty.2) of an ar-18800-03 driver shafts twisted at the tip while inserting a locking screw and no longer usable.This was discovered during the case with no patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user error of the device due to over-torquing the screw when inserting into the plate and/or hard bone.
|
| |
|