The device was not returned for analysis.A review of the production records and the corrective and preventive actions (capa) database was performed and revealed no indication of a product quality issue.Additionally, a query of the complaint handling database for the reported lot revealed no indication of a lot specific issue.The investigation was unable to determine a conclusive cause for the reported difficulties.It was reported that the microcatheter encountered resistance during advancement and removal over the guide wire.Factors that may contribute to difficulty advancing or removing a microcatheter over the guide wire include, but are not limited to, inner diameter of the balloon catheter, outer diameter of the guide wire, device support, buildup of procedural contaminants, challenging anatomy, user technique, and/or damage to the balloon catheter or guide wire.Based on the information provided and because the device was not returned for analysis, it is unknown what may have caused the reported difficulty during advancement or removal.Additionally, it is possible that manipulation of the guide wire, when resistance was met, contribute to the reported peeled core; however, this cannot be confirmed.There is no indication of a product quality issue with respect to manufacture, design, or labeling.The additional device referenced in b5 is filed under a separate report number.
|