| Model Number |
30001002001110012 |
| Medical Device Problem Code |
Defibrillation/Stimulation Problem (1573)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation for evaluation.The customer's report was observed and attributed to a fractured solder on a transformer on the pace/defib engine board.The pace/defib engine board will be replaced to remedy the report.The device will be recertified and returned to the customer.No trend is associated with reports of this type.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during a routine shift check by a clinician, the device displayed "defib fault 72" and "defib fault 76" messages.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|