Model Number GIF-XP170N
Medical Device Problem Code Failure to Clean Adequately (4048)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was observed during the device inspection, the nozzle, distal end, and bending section cover of the gastrointestinal videoscope had foreign objects.There were no reports of patient involvement.
 
Additional Manufacturer Narrative
This supplemental report is being submitted to provide the results of the manufacturer¿s investigation.A review of the device history record (dhr), a review of the instructions for use (ifu), as well as a historical trending analysis were conducted during this investigation.The review of the dhr did not find any abnormalities or anomalies identified during production.The device met all specifications upon release.The ifu contains the following information related to the failure mode: ¿the instruction manual operation manual ¿gif/cf-170 series, chapter 3 preparation and inspection¿ describes the methods for detection.The instruction manual reprocessing manual ¿gif/cf-170 series, chapter 5 reprocessing the endoscope (and related reprocessing accessories)¿ describes the methods for prevention.¿ based on the results of the investigation and the condition of the returned device, the specific foreign material found could not be identified.Additionally, while inspecting the subject device, there was no damage noted where the foreign material was discovered.Consequently, a definitive root cause for the reported failure mode could not be determined.Olympus will continue to monitor the field performance of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GASTROINTESTINAL VIDEOSCOPE
Common Device Name
GASTROINTESTINAL VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key20504455
Report Number9610595-2024-21545
Device Sequence Number8400934
Product Code FDS
UDI-Device Identifier04953170334146
UDI-Public04953170334146
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K123317
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 12/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberGIF-XP170N
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/15/2024
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/18/2024
Supplement Date Received by Manufacturer12/02/2024
Initial Report FDA Received Date10/22/2024
Supplement Report FDA Received Date12/03/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/10/2015
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-