| Model Number |
320574 |
| Medical Device Problem Code |
Failure to Deliver (2338)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Consumer reported needle clog during injection.Consumer does not prime, he does not re-use.Lot #: 3052557; catalog #: 320574; date of event: unknown; samples: available - sending mail kit.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction to: h6 (component code, investigation findings, and investigation conclusions) investigation summary: samples were received and an investigation was performed.This is the 3rd complaint for the reported lot number.A review of the manufacturing records was performed and no non-conformances were raised in association with this type of event for this lot.Embecta was able to duplicate or confirm the indicated issue as bent cannula npe.Based on trend analysis no further action is required at this time.Complaints received for this device and reported condition will continue to be tracked and trended.Based on the above, no additional investigation and corrective/preventative action (capa) or situational analysis (sa) is required.
|
| |
|
Search Alerts/Recalls
|