| Catalog Number |
XT22IS70NSC033N |
| Medical Device Problem Code |
Disconnection (1171)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/01/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the device swivels disconnects from the tracheostomy tube when removed.There was patient involvement, and no patient harm/adverse event reported.
|
| |
|
Additional Manufacturer Narrative
|
|
B3.Date of event: unknown.No information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: codes were updated.One device was returned for investigation.It was reported that the swivel/rotator has disconnected from the tracheostomy body.The device is sent with every icu medical tracheostomy.Based on the observed damage, a disconnect wedge was not utilized when disconnecting the hme, making the source of the defect user error.A device history record (dhr) review reported no discrepancies or non-conformance's during the manufacturing of the reported lot number.
|
| |
|