| Model Number |
N161 |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); Over-Sensing (1438)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Corrected fields: h6 impact codes.
|
| |
|
Event or Problem Description
|
|
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) exhibited oversensing of noisy signals.Technical services (ts) was consulted and discussed the stored episodes.No additional device issues were noted.This crt-d was explanted due to normal battery depletion.A new device was implanted.No adverse patient effects were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) exhibited oversensing of noisy signals.Technical services (ts) was consulted and discussed the stored episodes.No additional device issues were noted.This crt-d was explanted due to normal battery depletion.A new device was implanted.No adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned device was thoroughly inspected and analyzed upon receipt at our post market quality assurance laboratory.Visual examination of the device header and case noted no anomalies.Dimensional analysis of the header was completed.Each port measured as expected.The device was then exposed to simulated heart load conditions, and the defibrillation, pacing and sensing functions were tested.Impedance testing was completed and all measurements were within normal limits.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.Analysis did not identify any device characteristics that would have caused or contributed to the reported clinical observations.
|
| |
|
Event or Problem Description
|
|
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) exhibited oversensing of noisy signals.Technical services (ts) was consulted and discussed the stored episodes.No additional device issues were noted.This crt-d was explanted due to normal battery depletion.A new device was implanted.No adverse patient effects were reported.The device has been returned and analyzed.
|
| |
|