| Model Number |
10001000001110013 |
| Medical Device Problem Code |
Failure of Device to Self-Test (2937)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has not received the product for evaluation and this complaint is still under investigation.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during a routine shift check by a clinician, the device failed self test.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation for evaluation.The customer's report was duplicated and attributed to a faulty charging capacitor on the pace/defib engine board.The pace/defib engine board will be replaced to resolve the report.The device will be recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
|
| |
|