On 01-oct-2024, the following information was provided by the patient: v.A.C.® therapy was placed on hold for two weeks due to a fungal infection, allegedly related to the v.A.C.® drape.On 02-oct-2024, the following information was provided by the patient: on (b)(6) 2024, v.A.C.® therapy was removed due to an alleged fungal infection.Her skin was sweating and she is not sure if the irritation was caused by the v.A.C.® drape, or the v.A.C.® tubing rubbing against her skin.On 07-oct-2024, the following information was provided by the nurse: v.A.C.® therapy was placed on hold due to a periwound fungal infection that was treated with nystatin powder.There is no evidence of v.A.C.® tubing rubbing against patient¿s skin; photos display a fungal rash to the area covered by the v.A.C.® drape and there is documentation that a fungal odor was present.V.A.C.® therapy has remained on hold and fungal infection has resolved.On 22-oct-2024, a device history record review for v.A.C.® granufoam¿ dressing lot number a09429v005 and associated v.A.C.® drape was completed.All end release testing of product and packaging met specifications.The v.A.C.® granufoam¿ dressing was not returned; therefore, a device evaluation could not be performed.
|
|
Based on the information provided, it cannot be determined that the alleged fungal infection requiring medication is related to the v.A.C.® drape.All end release testing of product and packaging met specifications.A device evaluation could not be performed.Device labeling, available in print and online, states: warning: keep v.A.C.® therapy on: never leave a v.A.C.® dressing in place without active v.A.C.® therapy for more than two hours.If therapy is off for more than two hours, remove the old dressing and irrigate the wound.Either apply a new v.A.C.® dressing from an unopened sterile package and restart v.A.C.® therapy; or apply an alternate dressing, such as a wet to moist gauze, as approved during times of extreme need, by treating physician.Precautions: the v.A.C.® therapy system will not be effective in addressing complications associated with the following: ischemia to the incision or incision area, untreated or inadequately treated infection, inadequate hemostasis of the incision, cellulitis of the incision area.Dressing changes: wounds being treated with the v.A.C.® therapy system should be monitored on a regular basis.In a monitored, non-infected wound, v.A.C.® dressings should be changed every 48-72 hours, but no less than 3 times a week, with frequency adjusted by the clinician as appropriate.Infected wounds must be monitored often and very closely.For these wounds, dressings may need to be changed more often than 48-72 hours; the dressing changing intervals should be based on a continuing evaluation of the wound condition and the patient's clinical presentation, rather than a fixed schedule.Infected wounds: infected wounds should be monitored closely and may require more frequent dressing changes than noninfected wounds, dependent upon factors such as wound conditions, treatment goals.Refer to dressing application instructions (found in v.A.C.® dressing cartons) for details regarding dressing change frequency.As with any wound treatment, clinicians and patients / caregivers should frequently monitor the patient's wound, periwound tissue and exudate for signs of infection, worsening infection, or other complications.Some signs of infection are fever, tenderness, redness, swelling, itching, rash, increased warmth in the wound or periwound area, purulent discharge or strong odor.Infection can be serious, and can lead to complications such as pain, discomfort, fever, gangrene, toxic shock, septic shock and/or fatal injury.Some signs or complications of systemic infection are nausea, vomiting, diarrhea, headache, dizziness, fainting, sore throat with swelling of the mucus membranes, disorientation, high fever, refractory and/or orthostatic hypotension or erythroderma (a sunburn-like rash).If there are any signs of the onset of systemic infection or advancing infection at the wound site, contact the treating physician immediately to determine if v.A.C.® therapy should be discontinued.Disclaimer: this information is submitted pursuant to 21 cfr 803, in compliance with the medical device reporting requirement and should not be considered to be an admission that a kinetic concepts, inc.Product malfunctioned, is defective or has caused serious injury.
|