• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TANDEM DIABETES CARE T:SLIM INSULIN DELIVERY SYSTEM; INSULIN PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TANDEM DIABETES CARE T:SLIM INSULIN DELIVERY SYSTEM; INSULIN PUMP Back to Search Results
Model Number 004628
Medical Device Problem Code Failure to Sense (1559)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/03/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No product was returned for evaluation.Should new relevant information become available, a supplemental report will be submitted.H3 other text : device not returned.
 
Event or Problem Description
It was reported that the pump touch screen was unresponsive.There was no adverse impact to the customer¿s blood glucose level.The customer declined follow-up from tandem technical support regarding the issue, therefore no additional information was obtained.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
T:SLIM INSULIN DELIVERY SYSTEM
Common Device Name
INSULIN PUMP
Manufacturer (Section D)
TANDEM DIABETES CARE
12400 high bluff drive
san diego CA 92130
Manufacturer Contact
michael trier
san diego, CA 92130
8584011451
MDR Report Key20509055
Report Number3013756811-2024-200153
Device Sequence Number21108349
Product Code LZG
UDI-Device Identifier00852162004545
UDI-Public(01)00852162004545
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K111210
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 10/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number004628
Device Catalogue Number007262
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/03/2024
Initial Report FDA Received Date10/22/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/01/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient Age17 YR
Patient SexFemale
Patient EthnicityNon Hispanic
-
-