| Catalog Number |
0279351400 |
| Medical Device Problem Code |
Failure to Deliver Energy (1211)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/25/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the device had insufficient / no energy delivered.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that the device had insufficient / no energy delivered.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: the device does not coagulate, does not cut, buttons do not respond.The failure(s) identified in the investigation is consistent with the complaint record.The probable root cause/s could be a probe short due to a fluid ingress reaching the pcb board or, a broken/damaged bond of the suction tubing.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.
|
| |
|