| Catalog Number |
8250003 |
| Medical Device Problem Code |
Overheating of Device (1437)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
In this event it is reported that a cavitron touch, 240v-uk, (b)(6) that was used, it is alleged that the inserts get very hot.No injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Since an overheating insert could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation findings: failure mode: hot insert.Root cause: various cable problem: no contact or bad contact in the cable (wear or shock) / sheath of a damaged cable (wear or shock).Conclusion code: routine repair.
|
| |
|