STRYKER ORTHOPAEDICS-MAHWAH SIZE 8 ACCOLADE II 127 DEG; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
|
Back to Search Results |
|
| Catalog Number |
6721-0837 |
| Medical Device Problem Codes |
Corroded (1131); Degraded (1153)
|
| Health Effect - Clinical Codes |
Pain (1994); Metal Related Pathology (4530)
|
| Date of Event |
03/27/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Reported event: an event regarding abnormal ion level & corrosion involving a accolade stem was reported.The event was not confirmed.Method & results: product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.Clinician review: no medical records were received for review with a clinical consultant product history review: review of the product history records indicate devices were manufactured and accepted into final stock with no reported relevant discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: the exact cause of the event could not be determined because insufficient information was provided.Further information such as pathology reports and the revision operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.
|
| |
|
Event or Problem Description
|
|
It was reported by counsel that allegedly the patient underwent right tha on march 1, 2016, and was implanted with an lfit cocr femoral head.He was revised on march 27, 2024, due to alleged persistent pain, elevated metal ion levels and corrosion.
|
| |
|
Search Alerts/Recalls
|
|
|