| Model Number |
SUTR ANCH PEEK S-TAK KNOTLESS HIP USE |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/03/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Event or Problem Description
|
|
On 10/10/2024, it was reported by a sales representative via email that the suture from two ar-1938phs peek, knotless hip suturetak shred when passed around the labrum.This was discovered during a procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6.The complaint device was not received for evaluation.Based on the provided information, which may include the device (if available and returned), pictures, videos, event descriptions, and any additional field data, arthrex concluded that the most likely cause of the reported failure is user error.This includes pulling the sutures close to a rough bone.The complaint allegation was confirmed based on the customer's attached picture.
|
| |
|