|
BECTON, DICKINSON & CO. (BROKEN BOW) BD VACUTAINER® SODIUM FLUORIDE/POTASSIUM OXALATE 15MG/12MG; TUBES, VIALS, SYSTEMS, SERUM SEPARATORS, BLOOD COLLECTION
|
Back to Search Results |
|
| Catalog Number |
367925 |
| Medical Device Problem Code |
Coagulation in Device or Device Ingredient (1096)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/25/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that while using bd vacutainer® sodium fluoride/potassium oxalate 15mg/12mg, the customer had clotting issues on unspecified number of tubes.No patient impact reported.
|
| |
|
Additional Manufacturer Narrative
|
|
E1: initial reporter phone and email: (b)(6).Investigation summary: bd had not received samples or photos for investigation.Therefore, 100 retention samples from bd inventory were evaluated by visual examination and no gel issues were observed relating to clotting as all samples met specifications.Complaints for sample quality are under statistical control for the month of september 2024.At this time, further testing is not indicated.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint is unable to be confirmed for the indicated failure mode of clotting.Bd was not able to identify a root cause for the indicated failure mode.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends.Bd quality will continue to monitor sample quality complaints.
|
| |
|
Search Alerts/Recalls
|
|
|