• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. COMPREHENSIVE® REVERSE SHOULDER FIXED LOCKING SCREW; SHOULDER PROSTHESIS, REVERSE CONFIGURATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. COMPREHENSIVE® REVERSE SHOULDER FIXED LOCKING SCREW; SHOULDER PROSTHESIS, REVERSE CONFIGURATION Back to Search Results
Catalog Number 180553
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 09/26/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that approximately two and a half years (2.5) years following shoulder arthroplasty, the patient underwent a revision due to fractured baseplate screws.The baseplate, glenosphere, tray, bearing, and screws were removed and replaced with competitor products.The case was completed without any reported adverse outcomes to the patient.Attempts have been made, and no further information has been provided.
 
Additional Manufacturer Narrative
(b)(4).Multiple mdr reports were filed for this event.Associated reports are available for the applicable concomitant products below in the d10 narrative.D10: concomitant medical products, part (lot): 180554 (835070); 180554 (835070); 180553(161210); 115397 (100860).Associated product information: 110027734 (182270); 115313 (j7177943); 118000 (j7210693); 110031424 (65163892); - 110031402 (65255208).The customer has indicated that the product will not be returned to zimmer biomet for investigation per the hospital.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information is found that would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4; b5; d2; g1; g3; g6; h1; h2; h3; h6; h10 the reported event has been confirmed through the provided photo.Visual examination of the provided picture identified explanted devices with bio debris still on them.Screw fracture is able to be seen on multiple screws.However, as products did not return, further analysis could not be performed.A review of the device history records identified no deviations or anomalies during manufacturing.Medical records were not provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMPREHENSIVE® REVERSE SHOULDER FIXED LOCKING SCREW
Common Device Name
SHOULDER PROSTHESIS, REVERSE CONFIGURATION
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key20515497
Report Number0001825034-2024-02490
Device Sequence Number20972081
Product Code PHX
UDI-Device Identifier00880304677166
UDI-Public(01)00880304677166(17)320105(10)160810
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K130390
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/24/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue Number180553
Device Lot Number160810
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/27/2024
Supplement Date Received by Manufacturer01/21/2025
Initial Report FDA Received Date10/23/2024
Supplement Report FDA Received Date01/26/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/05/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
H11
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age72 YR
Patient SexMale
-
-