| Model Number |
MAJ-2113 |
| Medical Device Problem Code |
Connection Problem (2900)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Based on the results of the investigation, it is likely that the following led to the malfunction: external stress on the connecting tube caused by applying force in the wrong direction; however, a definitive root cause cannot be identified.Olympus will continue to monitor the field performance of this device.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
During the device evaluation, the connecting tube exhibited broken lock levers on the reprocessor side connector.There were no reports of patient involvement.
|
| |
|
Search Alerts/Recalls
|