MAUDE Adverse Event Report: STRYKER NEUROVASCULAR CORK TARGET TETRA 2.0MM X 4.5CM; DEVICE, NEUROVASCULAR EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER NEUROVASCULAR CORK TARGET TETRA 2.0MM X 4.5CM; DEVICE, NEUROVASCULAR EMBOLIZATION Back to Search Results
Catalog Number TG5412045
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Thrombosis/Thrombus (4440)
Date of Event 09/10/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during procedure once all implants had been detached and during the control runs which are done at the end of the procedure, clot formation was noted on the subject coil.The patient had to be given aggrastat infusion on the table which delayed the procedure.No further information is available.
 
Additional Manufacturer Narrative
H4 manufacturing date ¿ added.D4 expiration date - added.Due to the automated manufacturing execution system (mes) there are controls in the manufacturing process to ensure the product met specifications upon release.The subject device is not available; therefore, visual testing as well as functional testing cannot be performed.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported complaint could not be confirmed, and it could not be definitively determined if the device failed to meet specifications because the product was not returned.It was reported that during the procedure after implantation of the subject coil, clot formation was noticed on the subject coil once all implants had been detached and during the control runs which are done at the end of the procedure.The patient had to be given aggrastat infusion on the table which delayed the procedure.Based upon medical review, the event observed in this complaint is anticipated in nature as per the device risk assessment.As a product related root cause does not apply and the issue is due to a known physiological effect of the procedure and/or patient condition noted within the dfu, product labeling and/or risk documentation files, an assignable cause of anticipated procedural complication will be assigned to this complaint.
 
Event or Problem Description
It was reported that during procedure once all implants had been detached and during the control runs which are done at the end of the procedure, clot formation was noted on the subject coil.The patient had to be given aggrastat infusion on the table which delayed the procedure.No further information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TARGET TETRA 2.0MM X 4.5CM
Common Device Name
DEVICE, NEUROVASCULAR EMBOLIZATION
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI  NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI   NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key20517247
Report Number3008881809-2024-00493
Device Sequence Number20803516
Product Code HCG
UDI-Device Identifier07613327578591
UDI-Public(01)07613327578591(17)270128(10)24723169
Combination Product (Y/N)N
Initial Reporter CountrySF
PMA/510(K) Number
K222533
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberTG5412045
Device Lot Number24723169
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/02/2024
Supplement Date Received by Manufacturer10/28/2024
Initial Report FDA Received Date10/23/2024
Supplement Report FDA Received Date11/04/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/30/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient SexUnknown
-
-