This follow-up report is being submitted to relay additional information.The following sections were updated/corrected updated: b4, b5, g3, g6, h1, h2, h3, h6, h11 visual examination of the returned products found them to exhibit signs of repeated use and are missing components complaint was confirmed.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.Reported event is not related to a combination of products; therefore, a compatibility review is not applicable.Reported event did not occur in an operating room or as part of a medical procedure; medical records are not available for review.These devices are confirmed to have been a part of a previous design update.A field action notice was initiated to recommend against using ultrasonic cleaning as a sterilizing technique for these devices.These devices were subsequently re-designed to account for this failure mode with a new locking mechanism.It was determined that these devices were manufactured prior to this design change.Complaint is confirmed via product review.The root cause is considered to be a previously addressed design issue.No corrective actions, preventive actions, or field actions resulted after investigation of this event.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|