| Catalog Number |
80219 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/26/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H10: as the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that during an atherectomy and thrombectomy procedure, the catheter helix was allegedly fractured.The procedure was completed by using another device.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: manufacturing review: a manufacturing review was conducted and there was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: a physical investigation was performed for the catheter.During physical investigation the tube was kinked at two positions, at 6.5 cm and 30.5 cm from the tip of the catheter.The head part of the catheter was missing.The helix was found broken at 1.5 cm from the tip of the catheter.A clear root cause could not be established.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that during an atherectomy and thrombectomy procedure, the catheter helix was allegedly fractured.The procedure was completed by using another device.There was no reported patient injury.
|
| |
|
Event or Problem Description
|
|
It was reported that during an atherectomy and thrombectomy procedure, the catheter helix was allegedly fractured.The procedure was completed by using another device.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the rotarex products that are cleared in the us.The pro code and 510 k number for the rotarex products are identified in d2 and g4.H10: manufacturing review: a manufacturing review was conducted and there was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the physical sample was returned for evaluation, and a catheter was physically investigated.During physical investigation the tube was kinked at two positions, at 6.5 cm and 30.5 cm from the tip of the catheter.The head part of the catheter was missing.The helix was found broken at 1.5 cm from the tip of the catheter.A clear root cause could not be established.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H10: g3 h11: h3, h6 (conclusion) h11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|