Model Number 125270
Medical Device Problem Code Malposition of Device (2616)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
E1: complainant street address: (b)(6).Patient country: canada based on the available information, this event is deemed to be a reportable malfunction.Photograph, video and/or physical sample evaluation: there was a photograph associated with this case and in this, the reported defect can be seen.No unused return sample was expected.Batch record revision results: lot 3g05086 was manufactured 27/jul/2023, in mlk-3 line, with a total of (b)(4) market units (mkus).The complaints investigator performed a batch record review on 18/oct/2024, to verify if all the applicable procedures were followed and no issues were found; all the components for assembly were correct per bill of materials (bom) and all the tooling information documented was also correct, system application product (sap) material identification (id) 1051278 and manufacturing order (b)(4).Review of the batch record showed that all relevant tests required during the manufacturing process and final product release had been fulfilled and met the requirements.No discrepancy related to this issue were found within the documentation.Historical complaints review: on 18/oct/2024, complaints investigator ran a query in database in order to verify the complaints reported for the 3g05086 lot for the malfunction ¿skin barrier starter hole is defective (e.G misalignment or off center), leakage may occur (pre-cut only)¿ defect and no additional complaints were identified.Historical nonconformance review: on 18/oct/2024, complaints investigator ran a query in database looking for any in process nonconformance / corrective action / preventive actions (capa) (s) associated to the malfunction ¿skin barrier starter hole is defective (e.G misalignment or off center), leakage may occur (pre-cut only)¿ defect for the lot number 3g05086 and as result, no nonconformance / corrective action / preventive actions (capa) (s) for this malfunction were generated during the manufacturing process of the referenced lot.Defect rate analysis: there have only been 1 defective parts confirmed to date from a lot size 17500 products.This represents a defect rate of only (b)(4) which is well within an appropriate acceptable quality level (aql) for this defect which should 2.50% based on our process instruction (pi).In addition, all of the in-process testing on this lot did not find a single defective unit, which confirms that the lot is unlikely to breach an acceptable quality level (aql) of 2.50.This issue certainly appears to be an isolated incident, but more importantly to date, it is well within our accepted acceptable quality level (aql) level for this type of failure mode or defect.Conclusions: as a photograph was available for this complaint issue, it was evaluated and as a result, the reported malfunction for the observed product was confirmed, however, after performing the defect rate analysis for this issue, it was observed that the quantity of reportedly affected products is within the appropriate acceptable quality level (aql) for this defect.A batch record review was completed and showed that all relevant tests required during the manufacturing process and final product release had been fulfilled and met the requirements.No discrepancy related to this issue were found within the documentation.This issue will be monitored through the post market product monitoring review process, standard operating procedure (sop).To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.Fda registration number reporting site: 1049092 manufacturing site: 9618003.
 
Event or Problem Description
The end user reported that the starter hole was off centered for a skin barrier upon opening.The product was not used.A photograph depicting the issue was received from the complainant.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NATURA
Common Device Name
PROTECTOR, OSTOMY
Manufacturer (Section D)
CONVATEC DOMINICAN REPUBLIC INC
km 18.5 parque industrial
itabo, s.a. haina
san cristobal 91000
DR  91000
Manufacturer (Section G)
CONVATEC DOMINICAN REPUBLIC INC
km 18.5 parque industrial
itabo, s.a. haina
san cristobal 91000
DR   91000
Manufacturer Contact
pamela meadows
7815 national service road
suite 600
greensboro, NC 27409
3365424679
MDR Report Key20520868
Report Number9618003-2024-03917
Device Sequence Number20803502
Product Code EXE
UDI-Device Identifier00768455125173
UDI-Public00768455125173
Combination Product (Y/N)N
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Consumer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 10/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number125270
Device Lot Number3G05086
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/11/2024
Initial Report FDA Received Date10/24/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/27/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-