DEPUY SYNTHES PRODUCTS LLC KINCISE¿ PINNACLE® SHELL/LINER IMPACTOR; MOTOR, SURGICAL INSTRUMENT, AC-POWERED
|
Back to Search Results |
|
| Catalog Number |
1011-01-101 |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/16/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H11 additional narrative: the actual device has been returned and is currently pending evaluation.Once the evaluation has been completed, a supplemental medwatch report will be sent accordingly.
|
| |
|
Event or Problem Description
|
|
It was reported that during an unspecified surgical procedure, it was observed that the impactor device t-handle broke at the intersection of the t into three parts.It was not reported if there were any delays in the surgical procedure or whether a spare device was available for use.There was patient involvement.There were no injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the device t-handle was broken and failed pre-test for visual, end cap thread assessment and keyless fitting assessment.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to user error due to the damage to the t-handle which is consistent with the device having been dropped or excessive force used.
|
| |
|
Search Alerts/Recalls
|
|
|