• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. SHOWER SPRAY KIT; LAVAGE, JET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. SHOWER SPRAY KIT; LAVAGE, JET Back to Search Results
Catalog Number 00515049501
Medical Device Problem Codes Inability to Irrigate (1337); Failure to Power Up (1476)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/08/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This event is recorded with zimmer biomet under (b)(4).This medwatch is being filed as an initial / final report based on information discovered during the device evaluation.Functional testing found the device would not power on with the original battery pack but did power on and function normally in both modes when connected to a lab battery pack.Visual inspection of the interior of the pack found the batteries had leaked electrolyte inside of the pack.The batteries were installed in the correct orientation inside the pack.There did not appear to be damage to the wiring of the pack.No other damage was found.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.The root cause of the reported issue is attributed to manufacturing issue.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.G2: foreign : mexico.
 
Event or Problem Description
It was reported that during surgery, the pulse washing gun was opened ln which did not work when trying to wash the joint.The pulsatile washing didn't even turn on.There was no patient harm or injury.There was no medical intervention.There was a 15-minute surgical delay.During device evaluation and visual inspection, it was discovered that the interior of the pack found the batteries had leaked electrolyte inside of the pack.Due diligence is complete, there is no additional information available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHOWER SPRAY KIT
Common Device Name
LAVAGE, JET
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key20521902
Report Number0001526350-2024-01282
Device Sequence Number15433844
Product Code FQH
UDI-Device Identifier00889024375222
UDI-Public(01)00889024375222(17)260728(10)68809194
Combination Product (Y/N)N
Initial Reporter CountryMX
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/28/2026
Device Catalogue Number00515049501
Device Lot Number68809194
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/25/2024
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/22/2024
Initial Report FDA Received Date10/24/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/31/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age76 YR
Patient SexMale
-
-