| Model Number |
RAO100C |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Capsular Bag Tear (2639)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The reference (b)(4) has been allocated to this case by rayner.The event description provided states that the capsular bag ruptured during iol implantation.The rayner rayone preloaded injector risk analysis identifies the following as possible causes of "explosive expulsion of lens" resulting in capsule rupture; plunger advanced too quickly and forcing jammed plunger during iol insertion.Lens delivered in 's' orientation is also identified as a possible cause of capsule rupture within the risk analysis.Our review of production records for the rayone spheric rao100c batch 083222361 showed that all manufacturing and quality checks were conducted with successful results.A review of vigilance data confirms that this is an isolated event.No other incidents, of any type, have been received against the rayone spheric rao100c batch 083222361.There is insufficient evidence and information available to establish the cause of capsule rupture during iol implantation.
|
| |
|
Event or Problem Description
|
|
On 9th october 2024, rayner received notification from a us healthcare facility of an event that occurred during use of a rayone spheric rao100c.The event description provided states that the capsular bag ruptured during iol implantation.
|
| |
|