|
The reported event is confirmed cause unknown.Visual evaluation noted the sample was provided with the original bard inlay pouch and placed inside a large ziploc bag.The pigtail straighter was not provided.Visual requirements state to use "unaided eye at 12" to 18" distance under normal room lighting, unless otherwise noted." when evaluating the sample, the separation could be located on the proximal pigtail.The separation appears similar to what is seen when the material is cut due to no stretching or discoloration.When evaluating the suture, it seems it had been cut due to no stretching or discoloration.Dimensional: dimensional requirements per cd-7068-xx state the od to be 0.079" ± 0.002", guidewire to be 0.038" ± 0.001", tip id to be 0.043" ± 0.002" and tube id to be 0.050" +0.002" -0.001".The od of the sample was measured at 0.0781" and 0.0792" using a laser micrometer.The tip id was measured using a 0.043" pin gauge.The tube id where the separation occurred was measured using a 0.050" pin gauge.A 0.038" guidewire passed through the sample.Also received one photo sample that shows the top of stent on top of packaging with pigtail broken off.No actions can be taken at this time since a root cause was not identified.Dhr review did not show any problems or conditions that would have contributed to the reported event.The instructions for use were found adequate and states the following.Warnings: 1.Method for use: (1) when using the multi-length typestent, it should be avoided in the following cases.1)if you measure the length of patient¿s ureter and confirm that excessive coil part would be appear, consider using other stent which has different shape of tip and length.Ureteral stents with excessive coil parts have risks of knot formation at the tip of renal pelvis side during placement or removal.2) if any resistance is felt during removal, confirm the cause of the resistance with fluoroscopy and take remedial action to solve the problem.Excessive force during removal may lead to damage of the renal pelvis and/or ureter.(2) ureteroarterial fistula may be formed between the ureter and the aorta or the iliac artery and result in massive hemorrhage at the replacement of the ureteral stent if a ureteral stent is placed for a long term in a patient who has undergone the intrapelvic surgery or irradiation.Therefore, carefully monitor the condition of the patient, and in the event of acknowledging bleeding from the urethra, perform retrograde pyelography or angiography, and provide appropriate care.Contraindications: 1.Method for use: (1) do not reuse.(2) do not resterilize.2.Applicable patients.Do not use for the woman who is pregnant or may become pregnant.To avoid radiation exposure on pre-born baby from x-ray.Shape, configuration and principles, bard inlay optimatm stentset comprises the following components.1.Bard inlay optimatmureteral stent.The bard inlay optimatmureteral stent is a double pigtail ureteral stent with monofilament suture loop attached to aid in stent removal.The stent is available in two forms: a single size or a customized multi-length size.There are two types of stent: those which have side holes and have without side hole.Intended use & effect efficacy.Bard inlay optimatm stent set is indicated to relieve obstruction in a urinary tract and to be used for urethral catheterization from renal pelvis to urinary bladder.Each disposable kit contains several packaged medical devices that are necessary for stent placement in ureter.Directions for use: 1.Method of use: determine the proper stent length for the patient.This is generally calculated from the baseline pyelogram.Accurate measurements will optimize drainage efficiency and patient comfort.(1) insertion of the guidewire.1) remove the guidewire from the packaging, together with the guidewire holder.2) prior to removing the guidewire from the guidewire holder, inject sterile water through the port to activate the hydrophilic coating.3) remove the guidewire from the guidewire holder.Before using the guidewire, confirm the surface slides well when removing it from the holder.If you feel any resistance, do not force it, and inject the normal saline solution into the holder again, then try pulling it out once more.4) insert the guidewire either through the working channel of an endoscope or percutaneously, with the soft end of the guidewire first.5) advance the guidewire into the desired position until the tip is in the renal pelvis.Continuously confirm guidewire position either visually or under fluoroscopy.(2) nephroureterography.1) place the tigertailtmureteral catheter over the guidewire and insert it into the ureter.Note: in this case, it is also possible insert the ureteral catheter without using a guidewire, at the discretion of a physician.2) removing the guidewire.3) the ureteral catheter adapter is secured to the terminal end of the ureteral catheter, and performing the urine drainage or retrograde pyelography using with syringe.4) introduce the guidewire again for stent placement and remove the ureteral catheter from the cystoscope.(3) stent placement.1) in order to improve sliding, immerse the stent in a normal saline solution.2) confirm the guidewire position where it coils inside the renal pelvis.3) move the pigtail straightener over the proximal end (kidney coil end without the suture)of the ureteral stent, allowing easier insertion onto the guidewire.4) remove the pigtail straightener once the stent is secure on the guidewire and pass the stent over the guidewire through the cystoscope by using the pusher with radiopaque tip for proper placement.5) keep watching the distal end (bladder coil end with the suture) of the stent or radiopaque, proximal end of the pusher while advancing to proper position of the stent.Hold the guidewire to keep it from moving.6) stop advancing when the stent¿s distal end is identified.If the proximal end is inserted too much, pull the suture to modify the stent properly.7) confirm proximal end of the pusher and stent¿s distal end marker (bladder end), reaches the ureterovesical junction (uvj)by fluoroscopy.8) holding the stent in position with the pusher, cut the suture and pull it out.9) withdraw the guidewire.The stent will form a pigtail automatically.10) carefully remove the pusher.11) confirm position of the stent by fluoroscopy.Multi-length ureteral stent to accurately size this stent, count the marker bands as it is being advanced into the ureter.The first (wide) marker band on this device indicates 22 cm, followed by two 4 narrow bands at 24 cm and 26 cm.The last (wide) one indicates 28 cm.If you need to place the 30 cm or 32 cm lengths, use the attached suture or endoscopic forceps to gently pull back on the stent, unwinding the coil from the kidney.2.Precautions for use: ureteral catheter.(1) when using the tigertailtmureteral catheter, especially without the stabilizing support of a guidewire, be aware that a malfunction may occur where the flexible tip may detach if excessive force is made in contact with the walls of the bladder, ureter or renal pelvis.Should the flexible tip portion of the catheter become detached, retrieve with an endourology grasping device.(2) do not withdraw the ureteral catheter while it is deflected in endoscope.(3) avoid sharp bending of the ureteral catheter.(4) when performing drainage of urine, retrograde visualizations etc.Though the ureteral catheter, attach the ureteral catheter adapter to the tip of ureteral catheter.(5) do not over-tighten the catheter adapter.Over-tightening of the catheter adapter may occlude the lumen of the catheter stent.(1) do not forcibly insert or remove the stent.It may injure patient or/and damage this product.(2) avoid improper handling of stent such as bending, kinking, tearing, etc.Misuse could damage the overall integrity of the stent.(3) avoid contact with sharp edges as this may cause damage to the stent.If grasping device is used, the stent should be removed from ureter first.Tearing of the stent can be caused by sharp instruments.(4) determine the proper stent length for the patient.Selection of too short a stent may result in migration.(5) in the event of stent migration, cystoscopy or ureteroscopy should be used to return the stent to the original position or remove from the patient body.(6) any signs of infection in the location of the stent placement require removal of the stent.After checking the condition of the patient, a new stent should be placed.(7) care should be exercised when removing the stent to eliminate tearing or fragmentation.Guidewire: (1) do not forcibly insert or remove the guidewire.It may injure patient or/and damage the device.(2) do not manipulate, advance and/or withdraw the guidewire through a metal cannula or needle; to do so may result in damage the guidewire.Avoid contact with devices with sharp edges (such as metal dilator).(3) avoid using of the device when resistance is encountered (due to the size of a catheter, stent and/or working-channel of endoscope) as this may cause wear the guidewire coating.(4) do not use organic medical solutions or oily contrast medium on this device.These solutions may damage the device or decrease the lubricity of the device.(5) the guidewire is treated with a hydrophilic coating.Do not insert a stent over the device with its surface insufficiently wet.Never use dry gauze.Hydrophilic polymer coating can be damaged, increasing resistance when trying to insert catheter, stent or endoscope.(6) if unusual resistance is met during manipulation of the guidewire, do not force to 5 remove it.Carefully withdraw the guidewire as a unit.(7) do not use a retrieval device while the guidewire is in place; to do so may cause damage to the guidewire.(8) do not rub the guidewire with the edge of the holder.This could flake the hydrophilic coating.(9) avoid kinking, bending or twisting repeatedly the guidewire at acute bent site.It may lead to damage the guidewire.(10) never try to shape the guidewire.This could damage and break the cable core of the guidewire.(11) sufficient guidewire length must remain exposed to maintain a firm grip on the guidewire at all times.Precautions: 1.Important precautions: (1) the stent is not intended as a permanent indwelling device.It is recommended that the indwelling time not exceed 365 days.Periodic checks of the stent by cystoscopic and/or radiographic procedures are recommended at intervals deemed to be appropriate by the physician in consideration of the individual patient¿s condition and other patient specific factors.[all stents may be subject to varying degrees of encrustation when placed in the urinary tract.Encrustation may result in occlusion of the stent or pain or discomfort for the patient.] 2.Malfunction and adverse events: (1) malfunction, fragmentation, damage, guidewire kinking, difficulty in insertion, difficulty in removal, occlusion, migration, encrustation, difficulty in removal of the stent due to knotting of the coil.(2) adverse events , edema, loss of renal function, extravasation, pain/discomfort, fistula formation, perforation of kidney, renal pelvis, ureter and/or bladder, hemorrhage, peritonitis, hydronephrosis, infection, stone formation, urethral erosion, urethral reflux, separated piece remaining in body, urinary symptoms.Storage method and expiration date: 1.Storage: store in a dry, cool place away from heat, moisture, and direct sunlight.2.Expiration date.Indicated on the direct package and the outer box.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|