• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAILLEFER INSTRUMENTS HOLDING SARL WAVEONE GOLD SMALL 6-FILE STER 25MM; FILE, PULP CANAL, ENDODONTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAILLEFER INSTRUMENTS HOLDING SARL WAVEONE GOLD SMALL 6-FILE STER 25MM; FILE, PULP CANAL, ENDODONTIC Back to Search Results
Catalog Number A0756225G0S03
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 08/19/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As a result of this malfunction, the potential for surgical intervention exists to preclude permanent damage to a body structure or permanent impairment of a body function as evidenced by previous reported events with similar files.This event, therefore, is reportable per 21cfr part 803.The device was not returned for evaluation.However, the lot number was provided and retained-product testing and/or dhr review are planned.The results will be submitted as they become available.
 
Event or Problem Description
In this event it is reported that a waveone gold small 6-file ster 25mm broke during use.The broken part was not retrieved and was incorporated into the filling.No injury.
 
Additional Manufacturer Narrative
Investigation results summary: involved product that broke during use was not returned and cannot be analyzed.Moreover, no unused file is available for evaluation.Nothing unusual to report was found during dhr review (batch #1820873).Root causes are not identified.We will track this kind of event and monitor the trend.Potential root causes may be incorrect technique, overuse (mupltiple reuses may increase risks of breakage- number of use not formally known), excessive wear, or material issue (no analysis of the broken fragments possible).We moreover draw customer attention about the fact that waveone gold files are single use items.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WAVEONE GOLD SMALL 6-FILE STER 25MM
Common Device Name
FILE, PULP CANAL, ENDODONTIC
Manufacturer (Section D)
MAILLEFER INSTRUMENTS HOLDING SARL
ch. du verger 3
ballaigues vaud CH-13 38
SZ  CH-1338
Manufacturer (Section G)
MAILLEFER INSTRUMENTS HOLDING SARL
ch. du verger 3
ballaigues vaud CH-13 38
SZ   CH-1338
Manufacturer Contact
dan eagar
221 west philadelphia st.
york, PA 17401
7178457511
MDR Report Key20522398
Report Number8031010-2024-00128
Device Sequence Number18122787
Product Code EKS
UDI-Device IdentifierD716WGSMALL251
UDI-PublicD716WGSMALL251
Combination Product (Y/N)N
Initial Reporter CountryCH
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 12/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberA0756225G0S03
Device Lot Number1820873
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date10/21/2024
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/21/2024
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date10/24/2024
Supplement Report FDA Received Date12/05/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-