• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAILLEFER INSTRUMENTS HOLDING SARL PROTAPER GOLD F2 25MM STER; FILE, PULP CANAL, ENDODONTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAILLEFER INSTRUMENTS HOLDING SARL PROTAPER GOLD F2 25MM STER; FILE, PULP CANAL, ENDODONTIC Back to Search Results
Catalog Number A0411225G0203
Medical Device Problem Code Break (1069)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/10/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
There has been a previous report received where this malfunction resulted in a serious injury.Therefore, it must be presumed that recurrence of this malfunction could possibly cause or contribute to a serious injury or require medical or surgical intervention to preclude such.As such, this event is reportable per 21cfr part 803.The device was not returned for evaluation.However, the lot number was provided and retained-product testing and/or dhr review are planned.The results will be submitted as they become available.
 
Event or Problem Description
In this event it is reported that protaper gold f2 25mm ster broke during use.Further information requested.
 
Additional Manufacturer Narrative
Investigation results: summary: involved product that broke during use was not returned and cannot be analyzed.Moreover, no unused file is available for evaluation.Nothing unusual to report was found during dhr review (batch #1637278).Root causes are not identified.We will track this kind of event and monitor the trend.Potential root causes may be incorrect technique (motor setting not communicated), overuse (number of uses not communicated), excessive wear, or material issue (no analysis of the broken fragments possible).Additional information received: the broken part of the file has been removed from the patient's mouth and discarded.This is a follow up report for this additional information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROTAPER GOLD F2 25MM STER
Common Device Name
FILE, PULP CANAL, ENDODONTIC
Manufacturer (Section D)
MAILLEFER INSTRUMENTS HOLDING SARL
ch. du verger 3
ballaigues vaud CH-13 38
SZ  CH-1338
Manufacturer (Section G)
MAILLEFER INSTRUMENTS HOLDING SARL
ch. du verger 3
ballaigues vaud CH-13 38
SZ   CH-1338
Manufacturer Contact
dan eagar
221 west philadelphia st.
york, PA 17401
7178457511
MDR Report Key20522480
Report Number8031010-2024-00127
Device Sequence Number7983950
Product Code EKS
UDI-Device IdentifierD716PTGRF2251
UDI-PublicD716PTGRF2251
Combination Product (Y/N)N
Initial Reporter CountryCH
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 12/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberA0411225G0203
Device Lot Number1637278
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date10/18/2024
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/18/2024
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date10/24/2024
Supplement Report FDA Received Date12/05/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/01/2020
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-