SMITHS MEDICAL ASD, INC. CADD SOLIS HPCA PUMP; PUMP, INFUSION, PCA Back to Search Results
Model Number 2110
Medical Device Problem Code Insufficient Flow or Under Infusion (2182)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/01/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Month and year of event have been provided; day is unknown.One device was received for evaluation.The reported problem was verified by visual, accuracy and functional inspection.The cause of the device issue was an out of specification expulsor.Replaced the expulsor to correct the reported problem.The device passed all functional tests after repair.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.
 
Event or Problem Description
It was reported that the pump was under infusing.Occurred during testing.No patient involvement was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD SOLIS HPCA PUMP
Common Device Name
PUMP, INFUSION, PCA
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
Manufacturer Contact
reed covert
6000 nathan lane north
minneapolis, MN 55442
2247062300
MDR Report Key20523981
Report Number3012307300-2024-11936
Device Sequence Number8119507
Product Code MEA
UDI-Device Identifier15019517154887
UDI-Public(01)15019517154887(11)230323
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K170982
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 10/24/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2110
Device Catalogue Number21-2111-0402-78
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/26/2024
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/12/2024
Initial Report FDA Received Date10/24/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/23/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-