| Model Number |
2110 |
| Medical Device Problem Code |
Insufficient Flow or Under Infusion (2182)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/01/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Month and year of event have been provided; day is unknown.One device was received for evaluation.The reported problem was verified by visual, accuracy and functional inspection.The cause of the device issue was an out of specification expulsor.Replaced the expulsor to correct the reported problem.The device passed all functional tests after repair.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.
|
| |
|
Event or Problem Description
|
|
It was reported that the pump was under infusing.Occurred during testing.No patient involvement was reported.
|
| |
|