| Model Number |
DRIVER SHAFT, T7 HEXALOBE CANNULATED |
| Medical Device Problem Codes |
Break (1069); Defective Device (2588)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/08/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that according to the operating surgeon, there is a material defect in the screwdriver, because the device was used today and is now defective.No further information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Event or Problem Description
|
|
Update avoe 08-nov-2024.It was confirmed that during a metal-removal surgery one screwdriver broke off while the screw was being pulled out.Per complaint reporter there was no harm for patient, operator or third party.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.All metall parts could be removed from the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: b5, d9, g3, h3, h6.The complaint allegation is confirmed.Upon visual inspection, it was noted that the drive geometry of the distal tip was broken off.(damage) functional testing was not performed due to the damage to the device.It was not indicated if a torque-indicating adapter had been used with the device.The most likely cause of the reported and observed failure is user error.Specifically, the error involved applying excessive force, leading to over-torquing or over-engaging the driver within the screw head.
|
| |
|