| Model Number |
LF-DP |
| Medical Device Problem Code |
Peeled/Delaminated (1454)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was observed during the device inspection, that the subject device exhibited a peeled coating on the image tube.There were no reports of patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.It has been over two (2) years since the subject device was manufactured.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, and although we could not specify the root cause of the suggested event, it is likely the defect was caused by stress of repeated use, external factors, or handling.Olympus will continue to monitor field performance for this device.
|
| |
|