| Model Number |
Z800F |
| Medical Device Problem Codes |
Excess Flow or Over-Infusion (1311); Calibration Problem (2890); Infusion or Flow Problem (2964)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/07/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
There are previous complaints on the device.This pump underwent routine maintenance testing where it was detected the pump's performance fell outside the acceptable range for the flowrate %.Consequently, it necessitated an adjustment to the pump's flowrate %.It was confirmed the recalibration addressed the issue successfully.A capa has been opened in order to fully investigate and address the root cause of the reported event.Reference to complaint#: (b)(4).
|
| |
|
Event or Problem Description
|
|
On 10/07/2024, znyo medical received a report that during the preventive maintenance (pm), the device was reported to have a flow rate offset issue.Flow rate deviation too high (5.00).No report patient was involved.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon reassessment, the reported flow rate failure mode has been determined to be non-reportable.Events involving a faster flow rate are not reportable when flow rate accuracy is above +5% but below +15% specification.The severity of this failure is 1 - inconvenience or temporary discomfort.Our evaluation concluded that the event did not pose a risk to the health of patients, users, or bystanders.Furthermore, the report did not allege a serious injury or death, nor would a recurrence of the reported issue be expected to cause or contribute to a serious injury or death.Reference to complaint #: (b)(4).
|
| |
|
Search Alerts/Recalls
|