MENTOR TEXAS MENTOR SMOOTH ROUND MODERATE PROFILE; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE
|
Back to Search Results |
|
| Catalog Number |
3501620 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Codes |
Capsular Contracture (1761); Deformity/ Disfigurement (2360)
|
| Date of Event |
10/04/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: material rupture.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent breast augmentation revision with two 175cc mentor smooth round moderate profile saline breast prostheses.Post-operatively, the patient was diagnosed, via physical examination, with right breast implant deflation.As a result, the patient underwent breast implant removal surgery on (b)(6) 2024.
|
| |
|
Additional Manufacturer Narrative
|
|
On october 30, 2024, mentor received additional information indicating that the patient was also diagnosed with an extremely contracted right breast pocket, during the explantation surgery.The patient's implants were replaced bilaterally with the following devices: (right) 175cc mentor smooth round moderate profile catalog: 3501620, lot: 9983398, sn: (b)(6) and (left) 175cc mentor smooth round moderate profile catalog: 3501620, lot: 9983398, sn: (b)(6).
|
| |
|
Additional Manufacturer Narrative
|
|
On november 26, 2024, the mentor failure analysis lab received the device for evaluation.On november 27, 2024, the product investigation was completed.Device investigation summary: the product was returned to mentor for evaluation.Mentor conducted a visual inspection and microscopic examination of the returned device.Visual analysis of the returned sample revealed that the sal smooth rnd diap 175cc breast implant was found to have a tear on the posterior view measuring approximately 0.2 cm.In addition, a crease/fold was observed on the edges of the tear.A microscopic examination was performed and no instrument damage was observed at the edges of the tear.The evaluation determined that the possible cause of the rupture was consistent with normal wear.Deflation can occur at any time after implantation, but it is more likely to occur the longer the implant is implanted.Creating wrinkles or folds should be avoided during implantation or other procedures as it can result in deflation at a later time.Breast implants may also simply wear out over time.Mentor concluded that the capsular contracture in the patient´s breast was the result of the body´s individual physiological response to the implantation of a foreign object in soft tissue.Capsular contracture is a known complication associated with these devices and is referenced in our current product insert data sheet.The american society of plastic surgeons recommends and encourages member surgeons to always submit breast implants, capsule, and effusion to pathology for examination.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.No corrective and preventive action (capa) is required now.
|
| |
|