US SURGICAL PUERTO RICO AUTO SUTURE; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number OMS-T10BTNL
Medical Device Problem Codes Break (1069); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/10/2024
Type of Reportable Event Malfunction
Event or Problem Description
According to the reporter during a laparoscopic hernia repair, upon insertion and inflation of the balloon, after inserting the trocar, the balloon should be inflated but the air could not hold because of a hole in the balloon.In addition, the balloon moved from the trocar.This happened three times with one patient and twice with another patient.Another trocar was used to resolve the issue.There was no patient injury.
 
Additional Manufacturer Narrative
D10 concomitant products: oms-t10btnl, omst10btnl 10 blunt tip trocar w/o latex (lot#:p4d1023), oms-t10btnl, omst10btnl 10 blunt tip trocar w/o latex (lot#:p4d1023), oms-t10btnl, omst10btnl 10 blunt tip trocar w/o latex (lot#:p4d1023), oms-t10btnl, omst10btnl 10 blunt tip trocar w/o latex (lot#:p4d1023), oms-t10btnl, omst10btnl 10 blunt tip trocar w/o latex (lot#:p4d1023), oms-t10btnl, omst10btnl 10 blunt tip trocar w/o latex (lot#:p4d1023), oms-t10btnl, omst10btnl 10 blunt tip trocar w/o latex (lot#:p4d1023).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Additional information: d9, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the balloon was broken and disengaged from trocar sleeve.The reported issues could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTO SUTURE
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
Manufacturer (Section G)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key20526863
Report Number2647580-2024-04526
Device Sequence Number7983658
Product Code GCJ
UDI-Device Identifier10884521527386
UDI-Public10884521527386
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K924011
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/14/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberOMS-T10BTNL
Device Catalogue NumberOMS-T10BTNL
Device Lot NumberP4D1023
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/18/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/10/2024
Supplement Date Received by Manufacturer12/27/2024
Initial Report FDA Received Date10/24/2024
Supplement Report FDA Received Date01/14/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/16/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11