An end user reported an issue with an 8f low profile plastic vortex port detached poly catheter filled sh valved introducer.During a procedure, the physician inserted the port; however, the catheter detached and migrated into the patient's vessel.Ultimately, the port was removed; however, it is currently unknown how this event affected the patient.Despite multiple good faith effort attempts, no further details have been obtained.
|
|
The customer's reported complaint description of catheter detached was confirmed as it was returned in two pieces.The catheter tubing was separated at the ~26.8cm area.Break is clean and not jagged, looks to be cut with scalpel and is likely the intended junction of where catheter tubing was going to be connected to port housing or was connected to port housing and detached due to unsecure connection.Dimensional inspection of tubing and outlet tube confirmed it was in specification.Returned catheter and port was connected with a locking collar from stock with no issues observed making the connection.The root cause for this event could not be definitively determined but no manufacturing non-conformances observed.Device history record review of the packaging, assembly and component lots revealed no quality related issues or manufacturing deficiencies at the time of manufacture.Labeling review: the directions for use (dfu, 14600340-01) is provided in the port kit and contains the following directions and precautions: absence of a blood return or a poor blood return can be a sign of a potential complication such as occlusion, kinking, breakage, pinch-off syndrome, fibrin formation, thrombosis or malposition.This should be evaluated prior to device usage.· a blood return should be present prior to usage of device for any therapy or testing.· if the patient complains of pain, or there is swelling when the device is flushed or when medication or contrast media is administered, evaluate the device for infiltration, proper needle placement, and potential complications such as occlusion, kinking, breakage, pinch-off syndrome, thrombosis or malposition.Failure to assess these complaints or observations can lead to device failure.Precautions to avert device damage and/or patient injury during catheter placement: · carefully follow the catheter to port connection technique provided in the dfu to ensure proper device connection and to avoid catheter damage.· assure tight connection between port body and catheter.Potential complications use of an angiodynamics port system involves potential risks normally associated with the insertion or use of any implanted device or indwelling catheter including but not limited to: - air or catheter (or catheter fragments) embolism - catheter occlusion, malposition, dislodgement, fragmentation, migration, disconnection or rupture connecting the catheter a.Place catheter lock onto catheter.The catheter lock is bi-directional and can be mounted on the catheter in either direction.B.Trim the catheter to proper length at a 90° angle allowing sufficient slack to permit body movement, port connection and verify that the catheter is not kinked.C.Advance catheter over port stem to the midway point, slightly past the barb.D.Advance catheter lock until it engages with tactile and/or audible feedback.Note: to ensure proper assembly of port and catheter lock connector, a minimal gap (less than 0.5 mm) is expected.Note: if the catheter and locking collar are connected and then disconnected, the catheter proximal end must be re-cut to ensure a safe re-connection to the port.Precaution: prior to advancing the catheter lock connector, ensure that the catheter is properly positioned.A catheter not advanced to the proper region may not seat securely and lead to dislodgement and extravasation.The catheter must be straight with no sign of kinking.A slight pull on the catheter is sufficient to straighten it.Advancing the catheter lock over a kinked catheter may damage the catheter.A review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Reference (b)(4).
|