Model Number CT80LPPDVI
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 09/27/2024
Type of Reportable Event Serious Injury
Event or Problem Description
An end user reported an issue with an 8f low profile plastic vortex port detached poly catheter filled sh valved introducer.During a procedure, the physician inserted the port; however, the catheter detached and migrated into the patient's vessel.Ultimately, the port was removed; however, it is currently unknown how this event affected the patient.Despite multiple good faith effort attempts, no further details have been obtained.
 
Additional Manufacturer Narrative
The reported device has not yet been returned to the manufacturer for evaluation.An investigation into the root cause for the event is currently in progress.The results of the investigation will be sent via a follow up medwatch.Reference (b)(4).
 
Additional Manufacturer Narrative
The customer's reported complaint description of catheter detached was confirmed as it was returned in two pieces.The catheter tubing was separated at the ~26.8cm area.Break is clean and not jagged, looks to be cut with scalpel and is likely the intended junction of where catheter tubing was going to be connected to port housing or was connected to port housing and detached due to unsecure connection.Dimensional inspection of tubing and outlet tube confirmed it was in specification.Returned catheter and port was connected with a locking collar from stock with no issues observed making the connection.The root cause for this event could not be definitively determined but no manufacturing non-conformances observed.Device history record review of the packaging, assembly and component lots revealed no quality related issues or manufacturing deficiencies at the time of manufacture.Labeling review: the directions for use (dfu, 14600340-01) is provided in the port kit and contains the following directions and precautions: absence of a blood return or a poor blood return can be a sign of a potential complication such as occlusion, kinking, breakage, pinch-off syndrome, fibrin formation, thrombosis or malposition.This should be evaluated prior to device usage.· a blood return should be present prior to usage of device for any therapy or testing.· if the patient complains of pain, or there is swelling when the device is flushed or when medication or contrast media is administered, evaluate the device for infiltration, proper needle placement, and potential complications such as occlusion, kinking, breakage, pinch-off syndrome, thrombosis or malposition.Failure to assess these complaints or observations can lead to device failure.Precautions to avert device damage and/or patient injury during catheter placement: · carefully follow the catheter to port connection technique provided in the dfu to ensure proper device connection and to avoid catheter damage.· assure tight connection between port body and catheter.Potential complications use of an angiodynamics port system involves potential risks normally associated with the insertion or use of any implanted device or indwelling catheter including but not limited to: - air or catheter (or catheter fragments) embolism - catheter occlusion, malposition, dislodgement, fragmentation, migration, disconnection or rupture connecting the catheter a.Place catheter lock onto catheter.The catheter lock is bi-directional and can be mounted on the catheter in either direction.B.Trim the catheter to proper length at a 90° angle allowing sufficient slack to permit body movement, port connection and verify that the catheter is not kinked.C.Advance catheter over port stem to the midway point, slightly past the barb.D.Advance catheter lock until it engages with tactile and/or audible feedback.Note: to ensure proper assembly of port and catheter lock connector, a minimal gap (less than 0.5 mm) is expected.Note: if the catheter and locking collar are connected and then disconnected, the catheter proximal end must be re-cut to ensure a safe re-connection to the port.Precaution: prior to advancing the catheter lock connector, ensure that the catheter is properly positioned.A catheter not advanced to the proper region may not seat securely and lead to dislodgement and extravasation.The catheter must be straight with no sign of kinking.A slight pull on the catheter is sufficient to straighten it.Advancing the catheter lock over a kinked catheter may damage the catheter.A review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Reference (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOFLO VORTEX
Common Device Name
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
ANGIODYNAMICS, INC
10 glens falls technical park
glens falls NY 12801
Manufacturer (Section G)
ANGIODYNAMICS, INC
10 glens falls technical park
glens falls NY 12801
Manufacturer Contact
alexandra invencio
26 forest street
marlborough, MA 01752
5086587990
MDR Report Key20527352
Report Number1317056-2024-00202
Device Sequence Number20829053
Product Code LJT
UDI-Device IdentifierH787CT80LPPDVI0
UDI-PublicH787CT80LPPDVI0
Combination Product (Y/N)N
Initial Reporter StateDE
Initial Reporter CountryUS
PMA/510(K) Number
K190559
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 01/17/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCT80LPPDVI
Device Lot Number5825382
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/27/2024
Supplement Date Received by Manufacturer01/16/2024
Initial Report FDA Received Date10/24/2024
Supplement Report FDA Received Date01/17/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/25/2024
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient SexUnknown
-
-