H5 - labeled for single use?: correction.H8 - usage of device: correction.Manufacturer's investigation conclusion: the reported event of losses of connection to the heartmate touch app was not confirmed.Additional information provided stated that the losses of connection were due to a damaged power module patient cable.No product was returned for evaluation.Multiple attempts were made to obtain additional information from the customer regarding the event (including if the patient cable would be returned for evaluation); however, no response was received.The root cause of the reported event could not be conclusively determined through this analysis.Device history records could not be reviewed as the lot number of the product is unknown.Heartmate 3 instructions for use (ifu) chapter 4 ¿heartmate touch communication system¿ and the heartmate touch communication system quick start guide under ¿how to use the heartmate touch communication system¿ explain the operation of the heartmate touch system, including how to set up the system and connect the tablet to the wireless adapter and system controller, how to determine the status of the wireless adapter, and how to use the heartmate touch app.These documents also warn that the heartmate touch system may be interfered with by other equipment, even if that other equipment complies with cispr emission requirements.Heartmate 3 instructions for use (ifu) chapter 7 ¿alarms and troubleshooting¿ and the heartmate touch communication system quick start guide under ¿troubleshooting¿ cover all alarms (visual and audible), including the connection failed - heartmate touch app and the disconnected system controller - heartmate touch app messages, and the actions to take when the alarms occur.Heartmate 3 patient handbook section 6 "caring for the equipment" and heartmate 3 instructions for use (ifu) section 8 "equipment storage and care" describe how to care for and clean all equipment.Heartmate 3 patient handbook section 10 and heartmate 3 instructions for use (ifu) section f, both entitled ¿safety checklists¿, provide checklists to assist the patient in performing routine maintenance of heartmate 3 lvad.This section also informs the user to replace any equipment or system component that appears damaged or worn.No further information was provided.The manufacturer is closing the file on this event.
|