|
Device evaluation: during technical inspection the following was found: the front panel and the nozzle for the insufflation tube are deformed.Damage from a fall.The packaging in which the device was delivered has no visible damage, so it can be assumed that the damage was caused by the customer.The "gas volume reset" button for resetting the gas consumption to "0" does not provide any haptic feedback (this behavior is due to the damage to the front panel).However, resetting the co2 consumption is still possible using the button.However, it has been found in isolated cases that the device automatically jumps to the service menu when switched on.The reason for this is the pressure exerted on the "gas volume reset" button due to the damage to the front panel.The service menu is accessed by pressing the "gas volume reset" button + turning the power switch.After removing the cover, it can be seen that the deformed nozzle partially clamps off the silicone tube from the electronic pressure relief valve inside the device.Functional test: - the device works properly despite the damage to the front panel: overpressure behavior: correct.If there is overpressure in the abdomen, the pressure is released via the electric overpressure valve until the set pressure is reached again.In addition, an acoustic signal sounds via the buzzer if there is overpressure.O continuous test (settings: 12mmhg, 10l/min): no errors occurred during the continuous test.O 1h with abdomen without leakage: no errors detected.O 1h with abdomen and with various leaks that were generated with the following nozzles and connected directly to the abdomen (0.3mm, 0.6mm, 0.7mm, 1.0mm, 1.7mm): no errors detected.O 1h with abdomen without leakage: no error detected.O 1h with abdomen and with various leaks that were generated with the following nozzles and connected directly to the abdomen (0.3mm, 0.6mm, 0.7mm, 1.0mm, 1.7mm): no error detected.The pump control behavior shows abnormalities with a co2 flow of > 6l/min.A clicking noise can be heard from the linear actuator.The correct function of the device is not affected by this.Even if the linear actuator gets stuck, the device's safety function kicks in and excess pressure is detected and reduced.According to serialnumber-code "gf", the product was manufactured in july 2004.The product is no longer open for sale since 2019-05-17 and has reached the end of repair since 2023-10-31.Most likely root cause: an exact reason why the pressure in the patient rise from 12mmhg to 28mmhg could not be determined.Even if that were to happen, the safety mechanisms would kick in and the pressure would be released via the overpressure safety valve and an acoustic signal would be emitted.Further, the ifu 96116010, version 5.2 - 10/2018 states that it is the user's responsibility to make sure the unit operates properly before using it.If the deformation of the front panel and its nozzle were in the conditions we detected, then the ifu recommendations and warnings were ignored by the user.It could not be excluded, the cardiac arrest happened to the patient is caused by the insufflator unintended high pressure.It is ensured, the safety mechanism of the endoflator is working and therefore no functional damage of the unit is causing the event.The event is filed under internal karl storz complaint id: (b)(4).
|