| Medical Device Problem Codes |
No Apparent Adverse Event (3189); Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Codes |
Failure to Anastomose (1028); Hemorrhage/Blood Loss/Bleeding (1888); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/15/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during an unknown procedure, the clip applier is cutting the tissue when the jaw is closed, which is causing leakage and bleeding.Procedure delayed 20 minutes.A new clip applier was used to complete the procedure.No patient consequences reported.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 10/24/2024.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent: -did the device deliver a scissored clip which cut the structure? -did the device deliver a properly formed clip which cut the structure? -did the clip not feed into the jaws, resulting in the jaws cutting the structure upon firing? (if yes, please reference the ifu which states to ensure visualization of the clip in the jaws prior to firing.) -how was the leaking controlled? -how was the bleeding controlled? -was a blood transfusion required? if so how much was given? -is the current patient status known? -was there any patient consequence or change in the post-operative care of the patient as a result of the event? (extended hospital stay, readmission, re-operation, etc.)".Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 1/16/2025 d4: batch # unk investigation summary the product was returned to ethicon endo-surgery for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the lx107 device was received with no apparent damage.In an attempt to replicate the reported incident, the device was tested for functionality.Upon functional testing of the device, the instrument loaded, retained, and deployed 6 clips as intended.The instrument was fully functional and conforming.No damage was noted on the test skin, the jaws were visual examined and no sharp edged were noted.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Although no conclusion could be reached on the cause of the reported event, the instructions for use do contain the following caution: select the appropriate size ligaclip extra ligating clip cartridge and corresponding clip applier.With the cartridge slots facing away from the base, insert the cartridge into one of the channel openings of the lc800 stainless steel cartridge base.Ensure that the cartridge is past the channel opening and securely held in place.Grasp the applier in the box lock area using the pencil grip technique.Insert the jaws of the instrument into the individual cartridge slot, making sure the tips of the applier are perpendicular to the surface of the cartridge.Insert the applier until it stops.Do not force the applier.It should enter and withdraw from the cartridge smoothly.Keeping the applier jaws perpendicular to the surface of the cartridge, retract the applier from the cartridge.The clip will be securely held in the applier jaws.It is not necessary to maintain ring tension to hold the clip in the applier jaws.Position the clip around the tubular structure to be ligated.Apply sufficient force to fully close the applier to assure that the clip is satisfactorily placed and secured.The event described could not be confirmed as the device performed without any difficulties noted.An evaluation of the manufacturing record could not be performed as the required product identification number was not provided to complete the evaluation.
|
| |
|