| Model Number |
MINIMALLY INVASIVE BUNION CORR SYSTEM |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
10/14/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On 10/14/2024, it was reported by a sales representative via (b)(4) that an ar-8741c instrument case minimally invasive bunion correction systems driver broke.This occurred during a case causing extra or time to retrieve the broken piece, as well as creating larger incisions to do so (defeating the purpose of the procedure name).
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user-applied excessive force during the tightening process.
|
| |
|