Model Number 160002
Medical Device Problem Code Failure of Device to Self-Test (2937)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/04/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Hamilton medical ag comes to the conclusion: investigation ongoing.
 
Event or Problem Description
The following was reported to hamilton medical ag: detailed complaint and failure description: "check hardware compatibility" is shown.
 
Additional Manufacturer Narrative
Follow-up 1 (final) - additional information: the entry fields b4, g6, h2, h3, h6 and h11 have been updated.It was reported that a «hardware incompatibility fault» (as per user description) occurred on the device.No patient involvement and consequently no clinical consequences were reported.To address the described issue, the customer was provided with replacement components, including the inspiratory valve, ambient valve, and expiratory valve.No log files were submitted for analysis, and no further feedback was received confirming whether the replacement resolved the issue.Although the outcome could not be verified, it is assumed that the issue was resolved by replacing the valves, as no further issues have been reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HAMILTON-C2
Common Device Name
HAMILTON-C2
Manufacturer (Section D)
HAMILTON MEDICAL AG
via crusch 8
bonaduz, switzerland 7402
CH  7402
Manufacturer Contact
daniel elice
via crusch 8
bonaduz, switzerland 7402
SZ   7402
MDR Report Key20531035
Report Number3001421318-2024-02575
Device Sequence Number8324617
Product Code CBK
UDI-Device Identifier07630002800730
UDI-Public(01)07630002800730(11)150626(240)160002
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K121225
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/14/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number160002
Device Catalogue Number160002
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/07/2024
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date10/25/2024
Supplement Report FDA Received Date11/14/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/26/2015
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-